Testo differential pressure measurement tools in overview

 

 testo 6383

 testo 6381

 testo 6351

  testo_6383  testo_6381  testo_6351
Parameters

Differential pressure

Optional: Humidity / temperature

Differential pressure

Flow velocity

Volume flow

Optional:Humidity / temperature

Differential pressure

Flow velocity

Volume flow

Selectable measuring ranges

10 Pa to 10 hPa

10 Pa to 1000 hPa

50 Pa to 2000 hPa

Housing

Flat stainless steel housing for flush wall installation (panel design).

Metal housing

Plastic housing

Ethernet networking

- Integration of the transmitters into customers’ Ethernet network

- Integration of the transmitters into measurement data monitoring systems, e. g. testo SaverisTM

Area of application

Differential pressure monitoring between cleanrooms (optional: simultaneous measurement of temperature and humidity)

 

Monitoring positive and negative pressure in cleanrooms, operating theatres and isolation rooms

 

Additional monitoring of humidity and temperature in cleanrooms (optional)

Differential pressure monitoring between cleanrooms (optional: simultaneous measurement of temperature and humidity)

 

Differential pressure monitoring in filling processes and spraypainting systems

 

Monitoring drying processes

Differential pressure monitoring between cleanrooms

 

Differential pressure monitoring in filling processes in the process industry

 

Critical air conditioning technology (VAC systems)

Usual installation site in cleanroom

Critical zone: Surface flush installation in cleanroom wall

Normal zone or outside zone

Normal zone oroutside zone

Tendering documents
Technical Data

 

Integrate the new transmitters into testo SaverisTM

The measurement data monitoring system measures pressure, temperature and humidity values in the environment and in processes. find

In industrial processes, exact temperatures, humidity values and pressure relationships play a crucial role. 

In a multitude of applications, testo Saveris helps to collect, safely store and present these data wirelessly (analog coupler) or via Ethernet. A selection of alarms that can be used flexibly supports the responsible parties to keep the values in the required range.

 

Typical applications:

  • Monitoring ambient storage and production conditions
  • Monitoring humidity values, e. g. in air-conditioned cabinets
  • Monitoring temperatures, e. g. during heat treatment or in airconditioned cabinets
 

Calibration and validation/qualification services

Calibration:

 

Calibration service in laboratories and on site in numerous countries. Contact your local Testo subsidiary, testo industrial services or Testo’s sales partner in your country.

  

Validation / qualification

The Testo subsidiary testo industrial services (TIS) offers you customized validation and qualification for pharmaceutical projects:

  • DQ, IQ, OQ, PQ (qualification)
  • Mapping/distribution measurements (for optimum probe site determination)
  • Customized risk analysis incorporating GAMP5
  • Documentation optionally in Testo or customer format

  

Optimizing processes and saving time in commissioning and maintenance

The Testo transmitters are delivered ready to use. For professional application, the following functions are available via the easy-to-use software:

  • Parameterization of unit and scale
  • Sensor adjustment and adjustment of the analog outputs (humidity: 1-point, 2-point; pressure: n-point)
  • Parameterization and adjustment history of all activities of the P2A software are registered in the PC


An adjustment of the entire signal chain is possible directly at the measurement site thanks to the external interface. This saves time in commissioning and maintenance.

In addition to this, complete parameter files can be stored in the PC. The parameterization of spare transmitters or similar measuring points is thus possible with minimal time expenditure.

Contact
Testo AG
Testo-Strasse 1
79853 Lenzkirch
Germany

Telephone: +49 7653 681-700
Fax: +49 7653 681-701

E-Mail: info@testo.de
www.testo.com


Whether in cleanrooms, greyrooms, OP theatres or filling systems:

Lowest differential pressures must be maintained between different rooms or zones in order to prevent contaminated air from entering.

 

For this reason, continuous measurement and regulation of these low differential pressures (according to cleanroom norm ISO 14644: 5 - 20 Pa) is required. Annual proof (against zero potential and against adjacent rooms) of this must be provided according to ISO 14644.

 

Following data are missing

    Your Data:

     
     
     
     
    /
    /
     
    /
    /
     
     
     
     
     
    Please call me.
     
    Testo differential pressure
    Product brochure
     
     

    Additional data:

     
    Branch
     
    Industry

     
    Function
     
    How did you become aware of Testo?
     
    Yes, I woud like to receive regular information, and agree to receiving the free newsletters. I understand that I can cancel these at any time.
     
     
    * Mandatory fields
    Contact
    Testo AG
    Testo-Strasse 1
    79853 Lenzkirch
    Germany

    Telephone: +49 7653 681-700
    Fax: +49 7653 681-701

    E-Mail: info@testo.de
    www.testo.com